Abstract: TH-PO1287
A Decellularized Biological Vascular Graft for Hemodialysis Access: 18-Month First-in-Human Clinical Outcomes of LineMatrix
Session Information
- Late-Breaking Research Posters
October 22, 2026 | Location: Exhibit Hall A, Convention Center
Abstract Time: 10:00 AM - 12:00 PM
Category: Dialysis
- 803 Dialysis: Vascular Access
Authors
- Xuefeng, Qiu, Huazhong University of Science and Technology Tongji Medical College, Wuhan, Hubei, China
- Guo, Ying, Humatrix Medical Technology (Suzhou) Co., Ltd, Suzhou, Jiangsu, China
- Ye, Hong, Department of Nephrology, The Second Affiliated Hospital of Nanjing Medical University, Suzhou, Jiangsu, China
- Luo, Yuan, Department of Nephrology, The Second Affiliated Hospital of Nanjing Medical University, Suzhou, Jiangsu, China
- Chen, Shuangshuang, Humatrix Medical Technology (Suzhou) Co., Ltd, Suzhou, Jiangsu, China
- Li, Hua, Zhejiang University School of Medicine Sir Run Run Shaw Hospital, Hangzhou, Zhejiang, China
Background
Arteriovenous grafts (AVGs) are an important vascular access option for hemodialysis patients. Conventional ePTFE grafts have limitations including infection risk and limited patency. LineMatrix® is a decellularized bovine carotid artery graft developed as a biologic alternative. This first-in-human study evaluated its safety and performance.
Methods
This prospective, single-arm study (ChiCTR2500106717) enrolled 12 patients with end-stage kidney disease requiring hemodialysis. LineMatrix® grafts were implanted in the upper extremity and used for routine hemodialysis. Primary endpoint was primary patency at 3 months. Safety and patency outcomes were assessed through 18 months.
Results
Twelve LineMatrix® grafts were successfully implanted. Two patients were excluded due to non-device-related infection at day 29 and withdrawal of consent at day 89. Among the per-protocol population (n=10), primary patency rates at 3, 6, 12, and 18 months were 90.0%, 80.0%, 70.0%, and 46.7%, respectively. Five patients required a total of 12 interventions to maintain patency (corresponding to 0.72 interventions/patient-year). Cumulative patency was maintained at 100% throughout 18 months. Freedom from LineMatrix®-related infection remained 100% at 18 months despite repeated hemodialysis cannulation (2-3 cannulations/week; estimated cumulative cannulation number: 149-234). No device-related steal syndrome, cardiac complications, aneurysm, pseudoaneurysm, or seroma were observed.
Conclusion
LineMatrix® demonstrated favorable safety outcomes and durable cumulative patency through 18 months. A multicenter randomized controlled trial comparing LineMatrix® with ePTFE grafts is ongoing to further evaluate clinical outcomes (ChiCTR2500107160; n=326).
Acknowledgment
Commercial Support – Humatrix Medical Technology (Suzhou) Co., Ltd.
Figure 1. LineMatrix® graft implantation and clinical outcomes.
(A) Implantation of LineMatrix® graft. (B) LineMatrix® AVG after 18 months. (C) Doppler ultrasound measurements at 12-month follow-up. (D) Digital subtraction angiography (DSA) at the 3-month follow-up. (E) Primary and cumulative patency rates of LineMatrix® grafts in the PPS over 18 months follow-up.
Funding
- Private Foundation Support